The Noise Figure
Uncertainty Calculator
This calculator is designed
to determine the uncertainty of two-port devices like amplifiers or of
frequency converting devices like mixers and receivers.
Instructions - Overview
The Uncertainty
Calculator can be used for either of the following cases:
1. Modelling the
performance of your system. For this purpose, defaults are available for
Agilent's noise figure instruments and noise sources. These defaults have
typical1 values associated with them and can be useful for
estimating the effect individual parameters have in overall uncertainty levels.
However, actual equipment may have different values to those shown as the
default typical value. If you already own the equipment that you are modelling,
please refer to the calibration certificate for the measured instrument
uncertainties. With these actual values, the results you obtain will have
higher accuracy for your particular set up. For this operation please select
"User Defined" from the pull down selection menu.
(1 typical values were calculated using a wide range of noise source
calibration results)
2. Making actual
calculations of the uncertainty of your system. For this you will need to
obtain accurate values of all the associated parameters in question such as
match and gain. As mentioned above, for highest accuracy or for frequency
selective measurements, please consult the calibration certificates of your
instruments to obtain the measured uncertainty parameters of the equipment
being used.
Noise Source Selection
For the highest accuracy,
and therefore most meaningful results, the 'User Defined' option from the pull
down menu should be selected whenever the actual value of the noise source
calibration data is available. This allows the entry of the actual uncertainty
of the Excess Noise Ratio (ENR) and the 50Ω match (in dB, VSWR or
reflection coefficient), which can be for the noise source of any manufacturer.
Additionally, there are default values providing typical parameters for
standard Agilent Technologies noise sources.
Device Parameters
The noise figure and gain
values are the values that are read from the display on the measurement
instrument after a calibrated measurement is made. The DUT input and output
match can be obtained from the datasheet for the device or by measurement using
a network analyzer.
Noise Figure
Instrument Selection
For results
with the highest accuracy, as with the noise source selection, the 'User
Defined' option from the pull down menu should be selected whenever the actual
values for the noise figure analyzer/meter are available. This allows the entry
of the actual uncertainty of the instrument noise figure and gain. The values
of instrument noise figure and match should be obtained from calibration
reports; otherwise typical instrument default values are available. If values
for the NFA instrument noise figure and match versus frequency are required,
typical values can be found on the data sheet (5980-0164E - page 3 to 5)
available in the "Download" section of the Noise Figure Uncertainty
Overview page.
The user-defined option can
also be used for the noise source of any manufacturer. Additionally, there are
default values providing typical parameters for standard Agilent Technologies
noise figure analyzers/meters
Sweep
This function allows an
input parameter to be selected then sweep between two different values so that
the variation in uncertainty can be seen. The number of points in the sweep can
also be selected here.
Getting Results
Once all the parameters have
been entered the results can be viewed by selecting the 'Tabular Results'
button for single points or the 'Graphical Results' button for swept results.
These buttons are at the top of the display along with the 'Calculator' button.